Descripción
El Sarns 7000 Carretera de circunvalación Pulmonar de Corazón está disponible restaurado de DRE.
Especificaciones
Dimensions (with stand)
- Height: 46.75 inches (118.7 cm) maximum, 34.75 inches (88.3 cm) minimum
- Width: 22.25 inches (56.5 cm)
- Length: 40.25 inches (102 cm)
Power Requirements
- AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
Internal Battery
- Type: gel type, sealed and rechargeable.
- Voltage: 12 VDC nominal
- Quantity: 2
- Charging Time: two batteries take 20 minutes to recharge from fully discharged.
- Life expectancy: minimum 200 full charge/discharge cycles.
Alarms
- High Temp: Actuates at displayed incubator temperature of 39° ± 0.5 °C.
- Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
- Power Failure: Actuates if main AC power fails and no DC back-up power is present.
- Heater Temp: Actuates if heater temperature exceeds 77°C nominal.
- Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.
Temperature
- Set Point Range: 21.5°C ± 1.0°C to 38 °C
- Rise Time: 30 minutes
- Variability: 0.2°C
- Overshoot: 1.5°C maximum
- Uniformity: 1°C
- Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1°C
- Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5°C
- Observation Light Output: 10-foot candles at mattress level
-------------------------------------------------
The Sarns 7000 Heart Lung Bypass is available refurbished from DRE.