Reacondicionado - GE Marquette Mac 1200 ECG

Reacondicionado - GE Marquette Mac 1200 ECG

Web ID: 417

Condición: Refurbished

0 Los artículos están en tu carrito. Mirar Carrito ya.

¿Necesita asesoramiento sobre los equipos? ¡Pregunte a Nuestros Expertos!

(502) 244-6333

Descripción

El GE Marquette MAC 1200 máquina de ECG en reposo es una solución integral con características prácticas los médicos en los hospitales más pequeños, clínicas y consultorios médicos sabrán apreciar.

Especificaciones

Provides 10 seconds of instantaneous ECG acquisition and printing of any selected report format. Microcomputer-controlled 12 lead resting ECG system.

Physical Weight & Dimensions

  • Width: 14.5 inches (370 mm)
  • Height: 3.7 inches (94 mm)
  • Depth: 12.6 inches (320 mm)
  • Weight: 12.4 lbs (5.6 kg) with graphics display and battery

Power

  • Ratings: 10 - 120 V (0.32A) / 220 - 240 V (0.16A)
  • Frequency: 49 - 65 Hz
  • Battery type: NiCad, 18V, 1.4 Ah
  • Battery charge time: 4 hours

Signal Processing

  • Acquisition: 12 leads, simultaneously
  • Signal input: Type CF according to IEC
  • Digital sampling rate: 1000 samples/second/channel
  • Dynamic range: Differential signals for AC voltage ±, 10mV Superimposed DC voltage (Polarization voltage) ±600mV
  • Resolution: 5 uV
  • Frequency response:0.04 - 150 Hz
  • Common mode rejection:> 140 dB
  • Input impedance: > 100 M ohms
  • Leakage current:
  • Pacemaker detection sensitivity: All leads, duration 0.1 - 2.5 ms, amplitude > 5 mV

Display

  • Type: LCD 320 x 240 pixel, backlit, Contrast adjustable
  • Displayed data: Selected lead group, operation mode, lead check, heart rate, AC filter, muscle her, ADS, gain, speed, name, ID, 3 ECG waveforms.

Writer

  • Type: Thermal array print head
  • Resolution: Vertically 8 dots/mm horizontally 25um at 25 mm/s
  • Paper: 8.5 × 11 inches
  • Writer speed: 5, 25, 50 mm/s
  • Paper type: Z-fold

Keyboard

Type: Membrane keypad with tactile feedback / 46 keys

Special Functions

  • Anti Drift System (ADS): Automatic compensation of baseline fluctuations caused by polarization voltage fluctuations at the electrodes.
  • Baseline adjustment: ECG presentation on averaged baseline.
  • Electrode monitoring: Audible and visual indication on the LCD of disconnected electrodes or fine break. Each single electrode is monitored.
  • Copy: In the automatic 12 lead mode , after ECG, recording, copies of the ECG can be printed.
  • Test: Automatic performance test upon power up, including verification of the signal path starting at the signal input.